Signal dispersion measurements on the gas detector volume of a dual-energy multimedia digital imaging sensor
Loading...
Date
Journal Title
Journal ISSN
Volume Title
Publisher
IEEE
Abstract
In this study, experimental results on the signal dispersion of a gas detector volume of a multimedia imaging detector, operating on both gaseous and solid-state ionization principles, with specific emphasis on single X-ray exposure dual-energy radiography, are presented. The results of this study indicate that the gas detector volume of the multimedia detector technology exhibits excellent signal characteristics suitable for a large number of imaging applications. By focusing on the gas detector element, drift distance causes significant effects on both the amplitude and full-width at half maximum due to mobility dispersion and space-charge. An improved signal results with increasing gas pressure.
Keywords
Citation
Giakos, G. C., Chowdhury, S., Shah, N., Guntupalli, S., Vedantham, S., Suryanarayanan, S., ... & Orozco, A. (2003). Signal dispersion measurements on the gas detector volume of a dual-energy multimedia digital imaging sensor. IEEE Transactions on Instrumentation and Measurement, 52(5), 1566-1572.