6‐MV small field output factors: intra‐/intermachine comparison and implementation of TRS‐483 using various detectors and several linear accelerators

dc.contributor.authorGhazal, Mohammed
dc.contributor.authorWestermark, Mathias
dc.contributor.authorKaveckyte, Vaiva
dc.contributor.authorTedgren, Åsa Carlsson
dc.contributor.authorETAL:
dc.date.accessioned2022-02-14T05:47:03Z
dc.date.accessioned2023-08-19T08:17:40Z
dc.date.available2022-02-14T05:47:03Z
dc.date.available2023-08-19T08:17:40Z
dc.date.issued2019-11
dc.description.abstractThe TRS-483 Code of Practice (CoP) introduced and provided output correction factors to be applied to measured detector-reading ratios to obtain OFs for several small-field detectors. Detector readings for 0.5 cm × 0.5 cm to 8 cm × 8 cm fields were measured and normalized to that of 10 cm × 10 cm field giving the detector-reading ratios. Three silicon diodes, IBA PFD, IBA EFD (IBA, Schwarzenbruck, Germany), PTW T60017, and one microdiamond, PTW T60019 (PTW, Freiburg, Germany), were used. Output correction factors from the CoP were applied to measured detector-reading ratios. Measurements were performed on six Clinac and six TrueBeam linacs (Varian Medical Systems, Palo Alto, USA). An investigation of the relationship between the size of small fields and corresponding detector-reading ratio among the linacs was performed by measuring lateral dose profiles for 0.5 cm × 0.5 cm fields to determine the full width half maximum (FWHM). The relationship between the linacs' focal spot size and the small-field detector-reading ratio was investigated by measuring 10 cm × 10 cm lateral dose profiles and determining the penumbra width reflecting the focal spot size. Measurement geometry was as follows: gantry angle = 0°, collimator angle = 0°, source-to surface distance (SSD) = 90 cm, and depth in water = 10 cm.en_US
dc.identifier.citationGhazal, M., Westermark, M., Kaveckyte, V., Carlsson‐Tedgren, Å., & Benmakhlouf, H. (2019). 6‐MV small field output factors: intra‐/intermachine comparison and implementation of TRS‐483 using various detectors and several linear accelerators. Medical Physics, 46(11), 5350-5359.en_US
dc.identifier.doihttps://doi.org/10.1002/mp.13830
dc.identifier.urihttps://edms.wexl.in/handle/1/2631
dc.language.isoenen_US
dc.publisherAmerican Association of Physicists In Medicineen_US
dc.subjectMeasurementsen_US
dc.subjectDeterminationen_US
dc.subjectIntroduceden_US
dc.subjectAn investigationen_US
dc.title6‐MV small field output factors: intra‐/intermachine comparison and implementation of TRS‐483 using various detectors and several linear acceleratorsen_US
dc.title.alternativejournal Articalen_US
dc.title.alternativejournal Articalen_US
dc.typeArticleen_US