6‐MV small field output factors: intra‐/intermachine comparison and implementation of TRS‐483 using various detectors and several linear accelerators
Loading...
Date
Journal Title
Journal ISSN
Volume Title
Publisher
American Association of Physicists In Medicine
Abstract
The TRS-483 Code of Practice (CoP) introduced and provided output correction factors to be applied to measured detector-reading ratios to obtain OFs for several small-field detectors. Detector readings for 0.5 cm × 0.5 cm to 8 cm × 8 cm fields were measured and normalized to that of 10 cm × 10 cm field giving the detector-reading ratios. Three silicon diodes, IBA PFD, IBA EFD (IBA, Schwarzenbruck, Germany), PTW T60017, and one microdiamond, PTW T60019 (PTW, Freiburg, Germany), were used. Output correction factors from the CoP were applied to measured detector-reading ratios. Measurements were performed on six Clinac and six TrueBeam linacs (Varian Medical Systems, Palo Alto, USA). An investigation of the relationship between the size of small fields and corresponding detector-reading ratio among the linacs was performed by measuring lateral dose profiles for 0.5 cm × 0.5 cm fields to determine the full width half maximum (FWHM). The relationship between the linacs' focal spot size and the small-field detector-reading ratio was investigated by measuring 10 cm × 10 cm lateral dose profiles and determining the penumbra width reflecting the focal spot size. Measurement geometry was as follows: gantry angle = 0°, collimator angle = 0°, source-to surface distance (SSD) = 90 cm, and depth in water = 10 cm.
Keywords
Citation
Ghazal, M., Westermark, M., Kaveckyte, V., Carlsson‐Tedgren, Å., & Benmakhlouf, H. (2019). 6‐MV small field output factors: intra‐/intermachine comparison and implementation of TRS‐483 using various detectors and several linear accelerators. Medical Physics, 46(11), 5350-5359.
