Abstract
This study proposes an alternative fast calibration procedure for the estimation of the well-stirred condition of the reverberation chamber (RC) using coupling transfer gain function. The novel coupling transfer current gain function of RC environment is developed using the measured two-port S-parameters. The proposed procedure involves the determination of the frequency corresponding to the well-stirred operation ( fwso) of RC through the assessment of RC EM-field distribution using high order goodness of fit tests, and checking of the requirements of the independent samples for the particular stirred environment. The proposed procedure requires a portion of time (around 30 minutes) as compared to the conventional field uniformity long RC calibration procedure.
Keywords
Citation
Yousaf, J., Ghazal, M., Hasan, M. N., Lee, H., Nah, W., & Yang, J. G. (2020, July). Rapid alternative calibration procedure of reverberation chamber using coupling transfer gain function. In 2020 IEEE International Symposium on Antennas and Propagation and North American Radio Science Meeting (pp. 1273-1274). IEEE.