Investigation of Error Sources in Temperature Measurement Using Thermocouples in Water Impingement Cooling

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

Taylor & Francis

Abstract

The reliability of thermocouples with separation measuring junction in temperature measurement in the cooling process of hot steel plates with impingement jet has been investigated using direct and inverse finite element analysis (FEA). It is concluded that while the attachment of thermocouple wires on surface has negligible influence on surface temperature distribution during air cooling, the conduction of wires in a jet impingement water cooling process has significant effect on the measured temperature. The disturbance of the temperature field due to the introduction of a small hole for the installation of internal thermocouple has also been studied and showed similar but less pronounced effects to those of the surface measurement. An increased distortion of the temperature field is evident when the thermocouple is attached on the top surface directly above the bottom surface of hole.

Citation

Xu, F., & Gadala, M. S. (2005). Investigation of error sources in temperature measurement using thermocouples in water impingement cooling. Experimental heat transfer, 18(3), 153-177.

Endorsement

Review

Supplemented By

Referenced By